Overview
The Analog signal units allow for either a current or voltage analog output from the sensor.
Choose between non-submersible and submersible models:
| Depth Rating | Function | |
|---|---|---|
| DTU-A | Non-Submersible | DTU and Analog out functions - Current or Voltage out |
| ACU300 | 300m | Analog out functions - Current out |
| ACU1250 | 1250m | Analog out functions - Current out |
| AVU300 | 300m | Analog out functions - Voltage out |
| AVU1250 | 1250m | Analog out functions - Voltage out |
- The units can be terminated via SubConn or pigtail
- The units require power (6 - 40 V) and connection to the analog input of the telemetry unit
- The units have RS232 comms which can be used to validate and upgrade the firmware
Range
| Type | Range | pH | Step size | Example output | Resulting pH |
|---|---|---|---|---|---|
| Current Output | 4 - 20 mA | pH 2-10 | 0.5 pH/mA | 10.3 mA | pH 5.15 |
| Voltage Output | 0 - 5 V | pH 2-10 | 1.6 pH/V | 2.8 V | pH 4.48 |
note
The analog units require a 12 V / 250 mA power supply.
For further support please contact support@anbsensors.com