Command Line Prompt
This option provides access to additional sensor settings, as well as the commands to start, stop, abrade and reboot the sensor. Additionally, you can view the live data stream while the sensor is in operation.
Connecting to Serial Port: COM5 Baud Rate 57600
PCB Version: STM / IB
Sensor Type: AQ2_5m
Serial Number: 300239
Current Interface Firmware Version: IB_21.0
Current Driver Firmware Version: STM_21.0
Current Status: Idle
Main Menu - What would you like to do?
1 Measurement Settings
2 Install new Firmware
3 Command Line Prompt
4 Download CSV Data from Sensor
5 Backup CSV Data to ANB Cloud
Enter your choice:
If the sensor clock isn't set, utils will automatically set the clock to that of the host computer and the following message will be shown
Setting Sensor Clock.... Please wait
Once the clock is set the Command Menu will be displayed:
SENSOR COMMANDS
******************
START Start Scanning
STOP Stop Scanning
ABRADE Inform the sensor that the transducer has been abraded
REBOOT Save settings and restart the sensor
WAKE Access the sensor when in sleep
SETTINGS
******************
VERSION Display the sensor firmware revisions, serial number and time
TIME Change the date and/or time
BAUD Change the communications Baud Rate
MB_ADDRESS Change the system Modbus address
SDCARD Turn SD card on of off
IMMERSION Enable or disable the immersion sensor
AUTO_ABRADER Inform the sensor if you are using an automatic abrader
Type chosen command OR ctrl-c to exit to main menu.
******************
Immersion Command is not available on ATX sensors (S/N <300200)
START
Type start to tell the sensor to start scanning.
The following will be displayed:
Monitoring started.
Starting scan. Please wait for the live data string, or consider opening the GUI to view your live data in graphical format.
$ANB,OA39,2026:03:10:11:25:16,7.59,22.66,36.24,55.23,0,0,0,55,0000220,
Allow the sensor to run for approximately 30 minutes to reach measurement stabilisation to ensure accurate measurements. If you are updating or redeploying a legacy sensor, the stabilisation period may exceed 30 minutes.
Do not run the sensors in pH buffer solutions. They are for use in seawater - freshwater.
STOP
Type stop to save the current results and reset but leave the sensor powered.
The following will be displayed:
Monitoring stopped.
Type next command or Ctrl-C to return to the UTILS menu.
The updated splash screen will then be displayed
ABRADE
Type abrade to tell the sensor that the transducer has been abraded. This resets all the internal sensor settings
The following will be displayed:
Abrasion acknowledged.
Type next command or Ctrl-C to return to the UTILS menu.
REBOOT
Type reboot to save the current settings and restart the sensor so it is ready to run with the saved configurations.
The following will be displayed:
Rebooting sensor.
WAKE
Type wake to communicate to the sensor when in a sleep mode.
Allows you to communicate with the sensor when in autonomous scanning, interval delay, start delay, immersion detect delay or modbus mode
The following will be displayed:
Sensor setup functions are now available
If there is no activity for one minute, it will return to low-power mode, with the following message displayed:
The sensor is returning to its preprogrammed routine.
The wake command is only available in Utils. To wake the sensor and enable communication in serial terminal mode, send a # 7 0 0 command. When sending this command in terminal mode, allow a 50 ms delay between each character and send the carriage return only after all characters have been transmitted. This gives the sensor's processor sufficient time to decode the incoming command. If the sensor does not respond to the first attempt, repeat the command.
VERSION
Type version to display the sensors current configuration, along with the firmware revisions, the serial number and the current date and time.
The following will be displayed:
********************************************************************
ANB SENSORS
System Type : AQ2_5m
Interface Firmware Version : IB_20.0
Driver Firmware Version : STM_20.0
System Serial Number : 300239
System Date (MM/DD/YY) : 20/03/26
System Time (24 Hour) : 12:20
Current Status : Idle
Deployment Mode : Controlled
Deployment Interface : Terminal
Modbus address : 0x55
Power : Always Powered
Autonomous Start Delay : 0:00 Hours:Minutes
Interval Delay : 0 Minutes
Fast Profiling Mode : Disabled
Immersion Sensor : Enabled
SD Card : Enabled
********************************************************************
Should you be in discussions with ANB Sensors regarding the operation of your device, you may be asked to refer to this command.
TIME
Type time to change the internal clock and/or calendar.
After typing this command, the following is displayed:
Enter Date MM/DD/YYYY [ENTER]
>
Enter time (24 hour HH:MM) [ENTER]
>
On pressing [ENTER] the date and time is displayed:
Current date and time: 09/21/26 11:15:00
Type next command or Ctrl-C to return to the UTILS menu.
BAUD
Type baud to allow the user to change the baud rate to any one of:
115200, 57600, 56000, 38400, 28800, 19200, 14400, 9600
The following will be displayed:
Select baud rate:
1 = 9600
2 = 14400
3 = 19200
4 = 28800
5 = 38400
6 = 56000
7 = 57600
8 = 115200
Enter option [1-8]
>
Type Reboot to apply the change. Restart Utils with the correct baud rate.
MB_ADDRESS
Type mb_address to change the Modbus address.
The following will be displayed:
Enter Modbus address (0x01 - 0xFE):
>0x
Type next command or reboot to apply the change.
Ctrl-C to return to the UTILS menu.
SDCARD
The sensor has internal memory which can be switched off to save power if necessary. If the internal is switched off, no data will be saved in the sensor. All data must be saved externally.
Type sdcard to turn the sd card on or off
The following will be displayed:
Select SD card status: 1 = enabled, 2 = disabled
>
Type next command or Ctrl-C to return to the UTILS menu.
IMMERSION
- The immersion sensor checks whether the sensor is submerged in water.
- If the sensor is not immersed, it will stop measurement and set the Transducer Health status to 4.
- In Continuous Mode, the sensor will wait for 5 minutes before checking again.
- In Interval Delay Mode, the sensor will wait for the configured interval delay before rechecking.
- Measurement will automatically resume once immersion is detected.
Type immersion to enable or disable the immersion sensor.
This function is not available on ATX sensors (S/N <300200)
The following will be displayed:
Select Immersion status: 1 = enabled, 2 = disabled
>
Type next command or Ctrl-C to return to the UTILS menu.
When power cycled the immersion sensor defaults to enabled.
If immersion is disabled but the sensor is immersed and scanning, all data lines will be output, even the reference tracking measurement. The data will look like this:
2026-03-15T14:01:25.138615Z,$ANB,11E3,2000:00:00:00:06:09,99.99,18.27,99.99,99.99,5,0,99.99,0,0000000,
2026-03-15T14:01:38.976797Z,$ANB,FA03,2000:00:00:00:06:23,99.99,18.27,99.99,99.99,5,0,99.99,0,0000000,
2026-03-15T14:01:52.855003Z,$ANB,629E,2000:00:00:00:06:37,99.99,18.27,99.99,99.99,5,0,99.99,0,0000000,
2026-03-15T14:02:06.693033Z,$ANB,6318,2000:00:00:00:06:51,99.99,18.27,99.99,99.99,5,0,99.99,0,0000000,
2026-03-15T14:02:20.578871Z,$ANB,02EC,2000:00:00:00:07:05,99.99,18.27,99.99,99.99,5,0,99.99,0,0000000,
We recommend immersion being switched on in deployment conditions.
AUTO_ABRADER
Type auto_abrader to inform the sensor that you are using an automatic abrader.
After typing this command, the following is displayed:
Select auto-abrader status: 1 = enabled, 2 = disabled
>
Type next command or Ctrl-C to return to the UTILS menu.
You need specific firmware to run the auto-abrader. Please contact your ANB Sensors representative.
For further support please contact support@anbsensors.com