Skip to main content

Sensor Diagnostics & Status

Sensor diagnostics is a two digit number. The first digit gives information about the sensor's current status and the second digit gives information about any errors found in the sensor.

First DigitStatusExplanation
0SleepSensor is idle and/or following its pre-set interval routine
1Interval ScanningpH will be updated when measurement completes
2Continuous ScanningpH will be automatically updated as per our continuous sequence


Second DigitStatusAction
0Healthy SensorNone
1Clock Battery ErrorIf there is no external power to the sensor the real time clock will not hold the programmed time
- If the sensor is powered, the time set and data streamed, this failure is not an issue
- If the sensor is in automonous mode the clock will fail if disconnected from the power when it was programmed and placed on an external battery
-Users can record when the sensor was first switched on (where the sensor's time will be set to 0) and calculate the times externally, or, if this is not viable, contact support@anbsensors.com
2SD Card ErrorEither SD Card has been disabled or
A failing in the internal data save has occurred and no new data can be saved to the internal memory
- If the sensor is connected to an external communications system the sensor will continue to stream data, however no data will be saved in autonomous mode
- Please contact support@anbsensors.com
3System ErrorContact support@anbsensors.com