Introduction

This document describes how to perform the firmware update procedure.

To update the firmware, the system needs to have a RS232 interface. If your system currently runs with a RS485 interface, it needs to be converted. Instructions on how to do this can be found here.

ANB Sensors firmware update will always comprise the interface and driver firmware which has to be uploaded together.

Follow the procedures outlined below to ensure complete compatibility of the two systems.

Set-up

THE SENSOR SHOULD NOT BE COLLECTING DATA WHEN CONNECTED TO THE DATA RETRIEVAL UNIT

  1. Install Tera Term before using the S1000 https://osdn.net/projects/ttssh2/releases/
  2. Connect your S1000 to power and control system
  3. Launch Tera Term
  4. On Set-up menu select serial port from the drop down. In the resulting window select the correct com port
  5. Select speed = 19200, Data = 8bit, Parity = none, Stop bits = 1, flow control = xon/xoff
  6. Transmit delay set to 0 sec/char and 30 msec/line, click OK
  7. On Set-up menu select terminal
  8. New line options in resulting window should be Receive = CR+LF and Transmit = CR+LF, click OK
  9. Turn on CAPS LOCK
  10. Holding down the U key and ensuring the sensor is dry, power the sensor (Note: This should be done simultaneously)
  11. The monitor will display a version number and a ‘>’ symbol

Uploading Interface Firmware

  1. Release CAPS LOCK key and type pf [ENTER]
  2. The monitor will display ‘pf+’
  3. On tera term go to file > send file
  4. Browse to where firmware update has been saved and click ‘open’
  5. The update progress to the system is indicated by ‘.*.*.*.’ which will continue until the update is complete
  6. At completion the upload will have been successful if the last character displayed is a ‘+’ symbol
  7. If no ‘+’ sign is displayed, disconnect sensor from power and repeat Set-Up and Driver firmware update sections
  8. If no ‘+’ sign persists after repeating the procedures please contact your ANB Sensors representative
  9. Disconnect S1000 and prepare the sensor for the driver firmware update by repeating the ‘Set-Up’ procedure

Uploading Driver Firmware

  1. Release CAPS LOCK key and press t [ENTER]
  2. The monitor will display an instruction to press the U key.
  3. Turn CAPS LOCK on and press the U key
  4. The monitor will display a version number and a ‘>’ symbol
  5. Release CAPS LOCK key and type pf [ENTER]
  6. The monitor will display ‘pf+’
  7. On tera term go to file > send file
  8. Browse to where firmware update has been saved and click ‘open’
  9. The update progress to the system is indicated by ‘.*.*.*.’ which will continue until the update is complete
  10. At completion the upload will have been successful if the last character displayed is a ‘+’ symbol
  11. If no ‘+’ symbol is displayed, disconnect sensor and repeat Set-Up and Driver firmware update sections
  12. If no ‘+’ symbol persists after repeating the procedures please contact your ANB Sensors representative
  13. Power down the S1000 and then Power up
  14. If required, convert back to RS485. Instructions to do this can be found here

Restarting your sensor after firmware upgrade

  1. Reboot sensor
  2. Select MENU
  3. Select MODE
  4. Choose the mode in which you would like the sensor to run, as shown here
    The sensor will then run in the mode you selected

Please do not hesitate to contact us at support@anbsensors.com if you have any questions.