Setting up with a terminal program
We recommend lightly abrading the transducer prior to the first deployment
Connecting
The sensor is terminated using a MCBH6M connector with the following pin-out:
Pin | Function |
---|---|
1 | Ground |
2 | Power |
3 | RS232 Sensor Receive |
4 | RS232 Sensor Transmit |
5 | RS485 B - Inverting (for sensors with a serial number > 3000028) |
6 | RS485 A - Non-Inverting (for sensors with a serial number > 3000028) |
Because of the potential for different vendors to use different signal definitions, if you connect our sensors' A/B wiring to your communication device and communication does not work, try reversing the signals.
Communicating
- Connect sensor to power and control system
- Select correct com port
- Select speed = 57600, Data = 8bit, Parity = none, Stop bits = 1, flow control = none
- All lines should be terminated with a CR
Once set-up is complete, turn on the sensor. The sensor will beep once and the following will be displayed:
********************************************************************
ANB SENSORS
Sensor Type : OC300
Sub System Type : 300
Interface Firmware Version : I6.5
Driver Firmware Version : D6.0
System Serial Number : 3000029
System Date (MM/DD/YY) : 06/07/23
System Time (24 Hour) : 12:20
System Mode : Controlled Scanning
Scanning Interval : 0 Minutes
********************************************************************
If you would like the sensor to record the time please set the internal clock
Do not run the sensors in pH buffer solutions. They are for use in seawater - freshwater.
Whenever the sensor is to be powered off, the sensor must be shutdown properly, simply by typing 'shutdown', as shown here:
********************************************************************
ANB SENSORS
Sensor Type : OC300
Sub System Type : 300
Interface Firmware Version : I6.5
Driver Firmware Version : D6.0
System Serial Number : 3000029
System Date (MM/DD/YY) : 06/07/23
System Time (24 Hour) : 13:04
System Mode : Controlled Scanning
Scanning Interval : 0 Minutes
********************************************************************
shutdown
Shutting Down at > 2023:06:07:13:05:39
It is now safe to power down the system
If you are using a parsing script on a connected microprocessor to access the menu functions and change the parameters in the sensor you will need to write a script that issues commands mimicking accessing the Primary Menu.